CALIBRATION OF A WEIGHING LYSIMETER FOR MEASURING COCOA EVAPOTRANSPIRATION

ABSTRACT Well-operated and calibrated weighing lysimeters reliably determine the water demand of agricultural crops, and the quality of the data is evaluated through the error analyses attributed to the instrument. The objective of this study was to determine performance parameters of a weighing lysimeter and to evaluate its adequacy for the direct determination of cocoa crop evapotranspiration. The lysimeter with 1.44 m2 surface, supported on four load cells installed in the experimental area of the Graduate Program in Agricultural Engineering of the Universidade Federal do Recôncavo da Bahia, was calibrated at 3:00 a.m., using known masses at intervals of 02 minutes in five cycles of addition and removal. Data were collected using a datalogger coupled to four load cells with an individual capacity of 1000 kg, programmed to perform readings every 01 second, storing the averages of 15 minutes. The lysimeter showed excellent accuracy, with r2greater than 0.99 in the correlation between the standard masses and the signal of the load cells and standard error of linear regression estimate of 0.03 mm. The combination of repeatability and hysteresis errors with an uncertainty of 0.19% showed that the load cells have accuracy of ±0.07 mm, with resolution of 0.00033 mV, and sensitivity to detect mass changes below 0.1 mm, the system is adequate for direct measurement of cocoa ETc at time intervals of less than one hour.

Saved in:
Bibliographic Details
Main Authors: SILVA,GERLANGE SOARES DA, PEREIRA,FRANCISCO ADRIANO DE CARVALHO, SANTANA,RODRIGO ALMEIDA, MENESES,THAÍS NASCIMENTO, LOPES SOBRINHO,OSWALDO PALMA, ROSARIO,ALEX SANTANA DO
Format: Digital revista
Language:English
Published: Universidade Federal Rural do Semi-Árido 2020
Online Access:http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1983-21252020000300803
Tags: Add Tag
No Tags, Be the first to tag this record!