Spectrophotometric flow injection procedure to indirect determination of paracetamol in pharmaceutical formulations using o-tolidine as reagent

A spectrophotometric flow injection method for the determination of paracetamol in pharmaceutical formulations is proposed. The procedure was based on the oxidation of paracetamol by sodium hypochloride and the determination of the excess of this oxidant using o-tolidine dichloride as chromogenic reagent at 430 nm. The analytical curve was linear in the paracetamol concentration range from 8.50 x 10-6 to 2.51 x 10-4 mol L-1 with a detection limit of 5.0 x 10-6 mol L-1. The relative standard deviation was smaller than 1.2% for 1.20 x 10-4 mol L-1 paracetamol solution (n = 10). The results obtained for paracetamol in pharmaceutical formulations using the proposed flow injection method and those obtained using a USP Pharmacopoeia method are in agreement at the 95% confidence level.

Saved in:
Bibliographic Details
Main Authors: Fatibello-Filho,O., Vieira,H. J.
Format: Digital revista
Language:English
Published: Fundação Editora da Universidade Estadual Paulista Júlio de Mesquita Filho - UNESP 2008
Online Access:http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-46702008000200007
Tags: Add Tag
No Tags, Be the first to tag this record!